INSTRUCTIONS FOR USE
DUOPA (carbidopa and levodopa) enteral suspension
These instructions are for use along with any other instructions your healthcare provider gives you. Please read the Medication Guide before you start using DUOPA and each time you get a refill. For questions or problems, call DUOPA support toll free at 1-844-386-4968. |
 |
The CADD-Legacy® 1400 pump is used for delivery of DUOPA through a tube into your stomach attached to the longer straight (green) connector. Enteral nutrition should only be given by the shorter angled connector (white, blue, or violet) (see Figure A and Table. Connector Colors). |
Table. Connector Colors. |
Y-Connector Size |
Design |
Gastric (“g”) Port Color |
Intestinal (“i”) Port Color |
15 FR |
Original |
White |
|
|
New |
Blue |
Green |
20 FR |
Original |
White |
|
|
New |
Violet |
|
Note: The original design of the Y-Connector is represented in the Figures throughout this Instructions for Use.
This Instructions for Use provides information for the CADD-Legacy® model 1400 pump only. There are other CADD-Legacy® pump models available. Read the label on the back of the pump to make sure it is a model 1400 pump.
Your healthcare provider prescribed DUOPA for you. Your healthcare provider programs your prescription into the CADD-Legacy® 1400 pump. The CADD-Legacy® 1400 pump is approved for use with DUOPA. DUOPA is provided as medication inside cassettes that connect to the CADD-Legacy® 1400 pump.
The pump delivers DUOPA in 3 ways:
- Continuous Rate: Steady delivery of DUOPA delivered throughout the day while pump is on
- Morning Dose: A large dose of DUOPA given each morning
- Extra Dose: A small dose of DUOPA given as needed during the day
You will need the following items to complete these steps:
- Pump
- DUOPA cassette
- Coin, like a quarter
- Carrying bag
- Syringe
- Syringe connector
- Room temperature water
CADD-Legacy®-1400 Pump |
|
 |
 |
Display
The display shows programming information and messages. The main screen, which the pump displays most of the time, shows the following: |
 |
DUOPA Cassette The single-use DUOPA cassette is for use with the CADD-Legacy® 1400 pump. Battery Compartment Two AA batteries fit into the battery compartment. Cassette Latch The cassette latch secures the DUOPA cassette to the pump. |
WARNINGS and CAUTIONS
Failure to follow the Warnings and Cautions below could cause return of your symptoms, damage to the pump, serious injury, or may lead to death in rare cases.
WARNINGS
- Only use the pump in a manner described in this Instructions for Use, after you have received training by your healthcare provider.
- To avoid explosion hazard, do not use the pump near flammable explosive gases.
- Only use extension sets that are approved for use with DUOPA (See the Full Prescribing Information for DUOPA), pay attention to all warnings and cautions associated with their use.
- Always have new batteries available for replacement. If power is lost, DUOPA will not be delivered.
- If the pump is dropped or hit, the battery door or tabs may break. Do not use the pump if the battery door or tabs are damaged because the batteries will not be correctly secured. This may cause loss of power and DUOPA will not be delivered.
- If the pump is dropped or hit, look at the pump for damage. Do not use a pump that is damaged or is not functioning correctly.
- If a gap is present between the battery door and the pump housing, this means the door is not correctly latched. If the battery door becomes detached or loose, the batteries will not be correctly secured. This could cause loss of power and DUOPA will not be delivered.
- Use only DUOPA cassettes for pump accuracy and to make sure the pump works correctly. Attach the DUOPA cassette correctly. A detached or incorrectly attached DUOPA cassette could cause a problem with getting your DUOPA.
CAUTIONS
- Use only Smiths Medical accessories and replacement parts for the pump as using other brands may adversely affect the operation of the pump.
- Do not operate the pump at temperatures below 36°F (2°C) or above 104°F (40°C).
- Do not store the pump at temperatures below -4°F (-20°C) or above 140°F (60°C). Do not store the pump with a DUOPA cassette attached. Use the protective cassette provided when storing the pump.
- Do not keep the pump in humidity levels below 20% or above 90% relative humidity.
- Do not place the pump in cleaning fluid or water, or allow solution to soak into the pump, keypad, or battery compartment.
- Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners.
- Do not use rechargeable NiCd or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (heavy duty) batteries. They do not provide enough power for the pump to operate correctly.
- Do not store the pump for long periods of time with the batteries installed. Battery leakage could damage the pump.
Morning Procedure |
|
|
- Take the DUOPA carton containing the DUOPA cassettes out of the refrigerator. Check the expiration date on the carton. Do not use any of the cassettes if the expiration date has passed.
- Take a DUOPA cassette out of the carton. Return the carton with the remaining cassettes to the refrigerator. Do not use the cassette if the expiration date has passed or the cassette is damaged or empty. Leave the DUOPA cassette at room temperature for 20 minutes before using.
- Each DUOPA cassette may be used for up to 16 hours after removal from the refrigerator.
WARNING: Use only DUOPA cassettes to make sure the pump works correctly. |
1) Remove the cassette clip (see Figure B):
- Remove the cassette tube from its slot in the clip.
- Pull the clip from cassette to slide it off of the cassette top.
|
 |
2) Attach the DUOPA cassette to the pump (see Figure C):
- Hold the pump so the latch faces up.
- Hold the DUOPA cassette so the tube points down.
- Insert the DUOPA cassette hooks into the hinge pins at the base of the pump.
|
 |
3) Latch the DUOPA cassette into the pump:
- Hold the pump and DUOPA cassette upright against a flat surface.
- Press down on the pump, until the DUOPA cassette fits tightly against the pump (see Figure D).
- Use a coin to twist the latch counterclockwise until the latch lines up straight with the arrow (see Figure E).
|
 |
 |
WARNING: Attach the DUOPA cassette correctly. A detached or incorrectly attached cassette could cause a problem with getting your DUOPA. |
4) Remove the red cap on the end of the cassette tube (see Figure F). Save the red cap for use when you throw away the cassette. WARNING: Do not connect the red cap to the stomach tube. It will block DUOPA flow. |
 |
5) Connect the stomach tube to the cassette tube:
- While holding the stomach tube steady, twist off the white cap on the end of the longer straight (green) connector (see Figure G). WARNING: Do not twist the stomach tube.
- Connect the cassette tube to the end of the longer straight (green) connector (see Figure H). Do not connect to the shorter angled connector (white, blue, or violet).
|
 |
 |
6) Turn the pump on:
 until the display turns on.
- Wait approximately 30 seconds for the pump to review settings.
- Check for
 on the screen. PUMP STATUS: The pump is now on but not yet delivering DUOPA. |
|
|
7) Inspect the tubing for kinks or closed clamps. If needed straighten kinks or open clamps (see Figure I). |
 |
8) Start the pump:
 until 3 dashes appear and then disappear from the screen.
- Wait approximately 15 seconds for the pump to start running.
- Check for
 on the display. |
|
|
PUMP STATUS: The pump is now running. DUOPA delivery will begin as programmed by your healthcare provider. If the pump will not start, a message should appear on the display. Refer to the Alarms and Messages section. |
It will take between 10 minutes and 30 minutes to deliver your morning dose. To start delivery of your Morning Dose you will need to press the Morning Dose key 2 times. NOTE: If you are unable to deliver your Morning Dose, it may be too soon since the last Morning Dose to deliver another dose. You may need to wait longer. The time between Morning Doses is decided by your healthcare provider. |
9) The first key press shows the Morning Dose on the display.

 on the display. The number on your display is the Morning Dose of DUOPA your healthcare provider prescribed for you. |
|
|
10) The second key press starts Morning Dose delivery.
 a second time to deliver the Morning Dose.
 shows a countdown of your Morning Dose. |
|
|
PUMP STATUS: After the Morning Dose finishes, the pump will automatically begin delivering the Continuous Rate. RUN will appear on the display. This completes DUOPA delivery for your Morning Procedure. |
11) Insert pump into the carrying bag (see Figure J).
- Other carrying cases are also available. Refer to the specific Instructions for Use, which accompanies your carrying case.
|
 |
12) Wear the bag over your shoulder or neck:
- Place the bag strap over your shoulder or neck (see Figure K).
- Make sure the pump is in correct position (see Figure L).
|
 |
 |
Extra Dose |
1) Give an Extra Dose of DUOPA: NOTE: If you are unable to deliver the Extra Dose, it may be too soon since the last Extra Dose to deliver another and you may need to wait longer. The time between Extra Doses and the amount of DUOPA in the Extra Dose is decided by your healthcare provider.
 on the display.

- Listen for 2 beeps.
- The display will show
 |
PUMP STATUS: The pump is now delivering the Extra Dose. When it finishes, RUN will appear on the display and the Continuous Rate will continue to run. |
For instructions on changing a DUOPA cassette, see Changing the Cassette.
Evening Procedure |
|
|
You will need:
- 1 Syringe
- 1 Syringe connector
- Room temperature water
- 1 Coin, like a quarter
|
|
|
1) Remove the pump from the carrying bag (see Figure M). |
|
 |
2) Stop the Continuous Rate:
 until 3 dashes appear and then disappear from the display.
 on the display. |
|
|
3) Turn the pump off:
 until 3 sets of dots appear and then disappear from the display and the display turns off.
- Check that the display is off.
|
|
|
4) Clamp the cassette tube (see Figure N). |
 |
5) Disconnect the tubing:
- Twist the cassette tube to disconnect it from the longer straight (green) connector (see Figure O). WARNING: Do not twist the stomach tube.
- Replace the red cap on the cassette tube.
|
 |
6) Flush the longer straight (green) connector:
- Connect the syringe connector to the longer straight (green) connector.
- Fill a syringe with room temperature tap or drinking water. Do not use hot water as it could burn the wall of your stomach or intestine.
- Connect the syringe to the syringe connector (see Figure P). Do not over-tighten the syringe connector or it could break. Do not use the syringe connector if it is cracked or broken.
- Push the syringe plunger to flush the tube. Do not force the syringe if flushing the tube is difficult. Call your healthcare provider if you are unable or have difficulty flushing your tube.
- Remove the syringe and the syringe connector.
- Replace the white cap on the longer straight (green) connector (see Figure Q).
|
 |
 |
7) Flush the shorter angled connector (white, blue, or violet):
- Twist the white cap off the shorter angled connector (white, blue, or violet).
- Connect the syringe connector to the shorter angled connector (white, blue, or violet).
- Fill a syringe with room temperature tap or drinking water. Do not use hot water as it could burn the wall of your stomach or intestine.
- Connect the syringe to the syringe connector (see Figure R). Do not over-tighten the syringe connector or it could break. Do not use the syringe connector if cracked or broken.
- Push the syringe plunger to flush the tube.
- Remove the syringe and the syringe connector.
Replace the white cap on the shorter angled connector (white, blue, or violet) (see Figure S).
|
 |
 |
8) Remove the DUOPA cassette from the pump:
- Hold the pump and DUOPA cassette upright against a flat surface (see Figure T).
- Use a coin to twist the latch clockwise until the latch pops out (see Figure U).
- Remove the DUOPA cassette from the pump.
|
 |
 |
Changing the DUOPA Cassette |
- Take the DUOPA carton containing the DUOPA cassette out of the refrigerator. Check the expiration date on the carton. Do not use any of the cassettes if the expiration date has passed.
- Take a DUOPA cassette out of the carton. Return the carton with the remaining cassettes to the refrigerator. Do not use the cassette if the expiration date has passed or the cassette is damaged or empty. Leave the DUOPA cassette at room temperature for 20 minutes before using.
- Each DUOPA cassette may be used for up to 16 hours after removal from the refrigerator.
WARNING: Use only DUOPA cassettes to make sure the pump works correctly. |
1) Remove the pump from the carrying bag (see Figure V). |
 |
2) Stop the Continuous Rate:
 until 3 dashes appear and then disappear from the display.
 on the display. |
|
|
3) Turn the pump off:
 until 3 sets of dots appear and then disappear from the display and the display turns off.
- Check that the display is off.
|
|
|
4) Clamp the cassette tube (see Figure W). |
 |
5) Disconnect the tubing:
- Twist the cassette tube to disconnect it from the longer straight (green) connector (see Figure X). WARNING: Do not twist the stomach tube.
- Replace the red cap on the cassette tube.
|
 |
6) Remove the DUOPA cassette from the pump:
- Hold the pump and DUOPA cassette upright against a flat surface (see Figure Y).
- Use a coin to twist the latch clockwise until the latch pops out (see Figure Z).
- Remove the DUOPA cassette from the pump.
|
 |
 |
7) Remove the cassette clip on the new DUOPA cassette (see Figure AA):
- Remove the cassette tube from its secured slot in the clip.
- Pull the clip from the cassette to slide it off of the cassette top.
|
 |
8) Attach the new DUOPA cassette to the pump (see Figure BB):
- Hold the pump so that the latch faces up.
- Hold the DUOPA cassette so that the tube points down.
- Insert the DUOPA cassette hooks into the hinge pins at the base of the pump.
|
 |
9) Latch the new DUOPA cassette into the pump:
- Hold the pump and DUOPA cassette upright against a flat surface.
- Press down on the pump until the DUOPA cassette fits tightly against the pump (see Figure CC).
- Use a coin to twist the latch counterclockwise until the latch lines up straight with the arrow (see Figure DD).
WARNING: Attach the DUOPA cassette correctly. A detached or incorrectly attached cassette could cause a problem with getting your DUOPA. |
 |
 |
10) Remove the red cap on the end of the cassette tube (see Figure EE). Save the red cap to use when discarding the cassette. WARNING: Do not connect the red cap to the stomach tube as it will block DUOPA flow. |
 |
11) Connect the stomach tube to the cassette tube:
- While holding the stomach tube steady, twist off the white cap on the end of the longer straight (green) connector (see Figure FF). WARNING: Do not twist the stomach tube.
- Connect the cassette tube to the end of the longer straight (green) connector (see Figure GG). Do not connect to the shorter angled connector (white, blue, or violet).
|
 |
 |
12) Turn the pump on:
 until the display turns on.
- Wait approximately 30 seconds for the pump to review settings.
- Check for
 on the display. PUMP STATUS: The pump is now on but not delivering DUOPA. |
|
|
13) Inspect the tubing for kinks or closed clamps. If needed straighten kinks or open clamps (see Figure HH). |
 |
14) Start the pump:
 until 3 dashes appear and then disappear from the display.
- Wait approximately 15 seconds for pump to start running.
- Check for
 on the display. PUMP STATUS: The pump is now running. |
|
|
15) Insert the pump into the carrying bag (See Figure II). |
 |
16) Wear the bag on your shoulder or neck:
- Place the bag strap over your shoulder or neck (see Figure JJ).
- Make sure the pump is in correct position (see Figure KK).
|
 |
 |
Changing the Batteries: If you see LowBat or Battery Depleted on the display, change the batteries. Use 2 new AA alkaline batteries such as DURACELL® or EVEREADY® ENERGIZER®. The pump keeps all the important information when the batteries are removed.
WARNING:
- Always have new batteries available for replacement. If power is lost, DUOPA will not be delivered.
- If the pump is dropped or hit, the battery door or tabs may break. Do not use the pump if the battery door or tabs are damaged because the batteries will not be correctly secured. This may lead to loss of power and DUOPA will not be delivered.
- If a gap is present anywhere between the battery door and the pump housing, the door is not correctly latched. If the battery door becomes detached or loose, the batteries will not be correctly secured. This could cause loss of power and DUOPA will not be delivered.
CAUTION:
- Do not use rechargeable NiCd or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (“heavy duty”) batteries. They do not provide enough power for the pump to operate correctly.
- Do not store the pump for prolonged periods of time with the batteries installed. Battery leakage could damage the pump.
|
1) Ensure the pump is stopped. |
|
2) Push and hold the arrow button while sliding the battery door until it comes completely off the pump (see Figure LL). |
 |
3) Remove the used batteries (see Figure MM). |
 |
4) Install new batteries into the battery compartment. NOTE: Insert the batteries correctly based on the picture in the battery compartment. If you insert the batteries backwards, the display will remain blank. Reinsert the batteries, making sure to match the + and – markings with the battery compartment picture. |
5) Listen for a beep. PUMP STATUS: The pump is now powered. The power-up sequence will start, the pump will go through an electronic self-test, and then the pump will beep 6 times at the end of the power-up sequence. All of the display indicators, the software revision, and each setting will appear briefly. If you do not hear a beep, and the display is off, the pump is not powered. Check that the batteries are correctly inserted. |
6) Slide the battery door back onto the pump into its original closed position (see Figure NN). |
 |
Change the Morning Dose |
|
|
Your healthcare provider may have set your pump to allow for dose changes to your Morning Dose and Continuous Rate (Lock Level 1). Do not change your medicine dose without approval and training from your healthcare provider. Talk with your healthcare provider to decide when to change your Morning Dose and Continuous Rate. Do not change your Extra Dose unless your healthcare provider tells you to. If your Extra Dose requires changes, your healthcare provider will provide instructions. |
Change the Morning Dose |
WARNING: Do not use the Prime button. Priming is for use by your healthcare provider only. |
1) Turn the pump on:
 until the display turns on.
- Wait approximately 30 seconds for pump to review settings.
- Check for
 on the display.
PUMP STATUS: The pump is now on but not yet delivering DUOPA. |
|
|
2) Inspect the tubing for kinks or closed clamps. If needed, straighten kinks or open clamps (see Figure OO). |
 |
3) Start the pump:
 until 3 dashes appear and then disappear from the display.
- Wait approximately 15 seconds for pump to start running.
- Check for
 on the display. PUMP STATUS: The pump is now running. |
|
|
4) Change the Morning Dose: a. Press

- time.
b. Check for
 on the display. c. Press
 or
 to select the desired Morning Dose. d. Press
 to store the Morning Dose. e. Make sure you see the correct Morning Dose on the display. If not, repeat Steps 4c to 4e. |
|
|
5) Deliver the Morning Dose:
 1 time. NOTE: If you see “Value not saved” on the display, press NEXT and then repeat Steps 4c to 4e.
 shows a countdown of your Morning Dose. PUMP STATUS: After the Morning Dose finishes, the pump will begin delivering the Continuous Rate. RUN will appear on the display. NOTE: If you are unable to deliver a Morning Dose, it may be too soon since the last Morning Dose to deliver another and you may need to wait longer. The time between Morning Doses is decided by your healthcare provider. |
|
|
Change the Continuous Rate |
|
|
1) Stop the Continuous Rate:
 until 3 dashes appear and then disappear from the display.
 on the display. |
|
|
2) Change the Continuous Rate: a. Press
 2 times. b. Check for
 on the display. c. Press
 or
 to select the desired Continuous Rate. d. Press
 to store the Continuous Rate. e. Make sure you see the desired Continuous Rate on the display. If not, repeat Steps 2c to 2e. |
|
|
3) Start the pump:
 until 3 dashes appear and then disappear from the display. NOTE: If you see “Value not saved” on the display, press NEXT and then repeat Steps 2c to 2e.
- Wait approximately 15 seconds for the pump to start running.
- The display will show
 PUMP STATUS: The pump is now running. |
|
|
Alarms and Messages The table below shows some of the common alarms that you may hear from the pump. With all alarms, read the display before pressing
 to silence the alarm. |
What you see: |
What you hear: |
Meaning |
Response |
Error |
Two-Tone Alarm |
An error with the pump has occurred. |
Contact your healthcare provider. |
High Pressure |
Two-Tone Alarm |
There is pressure backed up in the tubing. |
Check tubing for clamps, kinks, or blockages. Make sure the red cap has been removed from the DUOPA cassette tube. Flush connectors if necessary. If it is not possible to flush the tubes, contact your healthcare provider as your tube may be blocked. |
LowBat |
3 Two-Tone Beeps Every 5 minutes |
The pump batteries are low. |
Change the batteries right away. |
Upstream Occlusion |
Two-Tone Alarm |
If your healthcare provider has the Upstream Occlusion Sensor set to ON and a blockage in the DUOPA cassette is detected, this alarm will sound. |
Detach the DUOPA cassette. Check if the DUOPA cassette is empty. If not empty, reattach the DUOPA cassette. Restart the pump to continue delivery. Contact your healthcare provider if the alarm continues. |
No message on display |
Two-Tone Alarm |
Batteries were removed within approximately 15 seconds after stopping the pump. |
Install new batteries to silence the alarm. Otherwise, the alarm will stop within a short period of time. |
Display shows current pump status |
2 Beeps (Long-Short) |
The DUOPA cassette is not lined up with the pump or DUOPA is not flowing from the DUOPA cassette to the pumping mechanism. Very cold or extremely thick DUOPA may cause this alarm as well. |
Press NEXT to silence the alarm. The pump continues to run. Make sure the DUOPA cassette is correctly lined up with the pump and DUOPA is flowing. Take the DUOPA cassette out of the refrigerator for 20 minutes before attaching to the pump. |
Battery Depleted |
Two-Tone Alarm |
Batteries are dead. |
Install new batteries. To continue delivery, restart the pump when completed. |
Key pressed, Please release |
Two-Tone Alarm |
Key is being held down. |
Stop pressing key. If the alarm persists, close the cassette tube clamp and remove the pump from use. Contact your healthcare provider. |
No Disposable, Clamp Tubing |
Two-Tone Alarm |
Disposable refers to the DUOPA Cassette. No Disposable means the DUOPA cassette was removed. The pump is not sensing proper cassette attachment. |
Clamp the cassette tube and disconnect it from your stomach tube. A DUOPA cassette must be correctly attached in order for the pump to run. Press NEXT to silence the alarm. |
No Disposable, Pump won't run |
Two-Tone Alarm |
Disposable refers to the DUOPA Cassette. You have tried to start the pump without a disposable DUOPA cassette attached. |
Press NEXT to silence the alarm. A DUOPA cassette must be correctly attached for the pump to run. |
Service Due See manual |
Two-Tone Alarm |
The pump is scheduled for service. |
Press NEXT to silence the alarm. The pump is still working, but contact your healthcare provider for further instructions. |
Frequently Asked Questions |
What if I drop the pump or hit it against a hard surface? Do the following right away:
- Check the DUOPA cassette latch on the side of the pump and make sure the line on the latch lines up with the arrow on the side of the pump.
- Gently twist, push, and pull on the DUOPA cassette to make sure it is still firmly attached.
- Check the battery door to make sure it is still firmly attached.
If the DUOPA cassette or the battery door is loose or damaged, do not use the pump. Stop the pump right away, close the tubing clamp, and contact your healthcare provider. |
What should I do if I drop the pump in water? If you accidentally drop the pump in water, pick it up quickly, dry it off with a towel, and call your healthcare provider. |
WARNING: If the pump is dropped or hit, look at the pump for damage. Do not use a pump that is damaged or is not working correctly. |
What should I do if I need to bathe while wearing the pump? You’ll need to detach the pump before you shower, bathe, or swim. Reattach the pump to the stomach tubing afterwards and restart it. |
What should I do if I need to have a medical test while wearing the pump? The pump may need to be removed prior to certain medical tests. Be sure to talk to your doctor about your DUOPA pump before you take these tests. |
STORAGE and DISPOSAL |
Storage
- Store DUOPA in the refrigerator with the temperature between 36°F to 46°F (2°C to 8°C).
- When the DUOPA cassette has been removed from the refrigerator, DUOPA should be used within 16 hours.
- The DUOPA cassettes are for single use only and should not be used for longer than 16 hours, even if some of the medicine remains. An opened cassette should not be re-used.
- Protect the cassette from light and keep it in the carton before using.
|
Throwing away your DUOPA cassette or batteries
- Throw away the DUOPA cassette as your healthcare provider tells you to.
- Throw away used batteries in a manner safe for the environment, and according to any regulations that apply.
|
This Instructions for Use has been approved by the U.S. Food and Drug Administration.
 |
AbbVie Inc. North Chicago, IL 60064, U.S.A. For DUOPA Support: 1-844-386-4968
|
|
Pump manufactured by: Smiths Medical ASD, Inc. 1265 Grey Fox Road St. Paul, MN 55112 USA Tel: 1-800-258-5361 www.smiths-medical.com |
|
|
03-B884 Revised: March 2019
Legacy 1400 Operator's Manual
CADD-Legacy® 1400 Pump
Model 1400
This online version differs from the printed version. Certain information that is not intended for patients has been removed.
This operator’s manual is for clinician use only. Read the entire operator’s manual before operating the pump.
This manual pertains only to the CADD-Legacy® 1400 pump. There are other CADD-Legacy® pump models available; review the rear label of the pump to ensure it is a CADD-Legacy® 1400 pump before programming. This pump is designed for enteral delivery of medication and can be programmed to deliver a continuous rate, a morning dose, and extra doses.
This manual is intended for clinician use only. Do not permit patients to have access to this manual. The pump has 3 security levels designed to limit patient access. Do not disclose the pump’s security codes or any other information that would allow inappropriate access to programming and operating functions.
The issue date of this operator’s manual is included on the back cover for the clinician’s information. In the event one year has elapsed between the issue date and product use, the clinician should contact Smiths Medical to see if a later revision of this manual is available.
Technical Assistance
If you have comments or questions concerning the operation of the CADD-Legacy® 1400 pump, please call the number given below. When calling, please specify the pump’s software revision. This information is located on the pump’s display during power up.
Our staff at Smiths Medical is available to help clinicians 24 hours a day with the programming and operation of the CADD-Legacy® 1400 pump.
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul, MN 55112 USA
Tel: 1 800 258 5361 (USA)
Tel: +1 614 210 7300
www.smiths-medical.com
Read this entire operator’s manual before operating the CADD-Legacy® 1400 pump.
Failure to follow the warnings and cautions below could result in return of symptoms, damage to the pump, serious injury, or death in extreme cases.
Please refer to the full prescribing information for DUOPA (carbidopa and levodopa) enteral suspension for indications and usage, contraindications, warnings, precautions, and adverse reactions.
Warnings
- This operator’s manual should be used by clinicians only. Do not permit patients to have access to this manual, as the information contained would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment.
- To avoid explosion hazard, do not use the pump in the presence of flammable anesthetics or explosive gases.
- The CADD-Legacy® 1400 pump and medication cassette reservoir are designed for enteral delivery of medication only. They are NOT intended for IV or other parenteral routes of infusion.
- Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment.
- Always have new batteries available for replacement. If power is lost, non-delivery of medication will occur.
- If the pump is dropped or hit, the battery door or tabs may break. Do not use the pump if the battery door or tabs are damaged because the batteries will not be properly secured; this may result in loss of power and non-delivery of medication.
- If a gap is present anywhere between the battery door and the pump housing, the door is not properly latched. If the battery door becomes detached or loose, the batteries will not be properly secured; this could result in loss of power and non-delivery of medication.
- System delivery inaccuracies beyond the stated accuracy may occur as a result of back pressure or fluid resistance, which depends upon temperature, medication viscosity, catheter size, extension set tubing, flow rate, and orientation of the pump system.
- Programming the pump at a delivery rate other than what is prescribed will cause over- or under-delivery of medication.
- Clamp the fluid path tubing and/or disconnect the tubing from the enteral access device before removing the medication cassette reservoir from the pump to prevent unintended delivery of medication.
- Use only approved DUOPA medication cassette reservoirs to maintain pump accuracy and assure proper pump operations.
- Use only extension sets approved for use with DUOPA, paying particular attention to all warnings and cautions associated with their use.
- Attach the cassette properly. The cassette is the part of the medication cassette reservoir that attaches to the pump. A detached or improperly attached cassette could result in unintended delivery of medication.
- Do not prime the fluid path with the tubing connected to a patient as this could result in over-delivery of medication.
- If the pump is dropped or hit, inspect the pump for damage. Do not use a pump that is damaged or is not functioning properly, as this could compromise patient treatment.
- The use of power supplies and a remote dose cord other than those listed in the electromagnetic emissions declaration may result in increased emissions or decreased immunity of the pump.
- The pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the user should verify normal operation of the pump in the configuration in which it is to be used.
- There are potential health hazards associated with improper disposal of batteries, electronics, and contaminated (used) reservoirs and extension sets. Dispose of used batteries, reservoirs, extension sets and other used accessories, or a pump that has reached the end of its useful life, in an environmentally safe manner, and according to any regulations that may apply.
Cautions
- Use only Smiths Medical accessories and replacement parts, as using other brands may adversely affect the operation of the pump.
- Do not operate the pump at temperatures below 2°C (36°F) or above 40°C (104°F).
- Do not store the pump at temperatures below -20°C (-4°F) or above 60°C (140°F). Do not store the pump with the medication cassette reservoir attached. Use the protective cassette provided.
- Do not expose the pump to humidity levels below 20% or above 90% relative humidity.
- Do not sterilize the pump or medication cassette reservoir as this could cause damage.
- When the upstream occlusion sensor is turned off, the pump will not detect occlusions in the medication cassette reservoir. Periodically inspect the medication cassette reservoir for decreasing volume. Undetected occlusions could result in under- or non-delivery of medication.
- Do not use rechargeable NiCd or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (“heavy duty”) batteries. They do not provide sufficient power for the pump to operate properly.
- Do not store the pump for prolonged periods of time with the batteries installed. Battery leakage could damage the pump.
- Prior to starting medication delivery, inspect the fluid path for kinks, a closed clamp, or other obstruction. An undetected occlusion may result in under- or non-delivery of medication and/or nuisance alarms.
- When you enter a new value, any lockout time already in effect will be cleared. An extra dose or morning dose could be requested and delivered immediately upon starting the pump, which may result in over-delivery of medication.
- Do not immerse the pump in cleaning fluid or water. Do not allow solution to soak into the pump, accumulate on the keypad, or enter the battery compartment. Moisture buildup inside the pump may damage the pump.
- Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners, as damage to the pump may occur.
- Do not expose the pump to therapeutic levels of ionizing radiation as permanent damage to the pump’s electronic circuitry may occur. The best procedure to follow is to remove the pump from the patient during therapeutic radiation sessions. If the pump must remain in the vicinity during a therapy session, it should be shielded, and its ability to function properly should be confirmed following treatment.
- Do not expose the pump directly to ultrasound, as permanent damage to the electronic circuitry may occur.
- Do not use the pump in the vicinity of magnetic resonance imaging (MRI) equipment as magnetic fields may adversely affect the operation of the pump. Remove the pump from the patient during MRI procedures and keep it at a safe distance from magnetic energy.
- This pump may interfere with ECG equipment. Monitor ECG equipment carefully when using this pump.
- CADD-Legacy® 1400 pumps are sealed units. A broken or damaged seal will, therefore, be considered conclusive evidence that the pump has been misused and/or altered, which voids any and all warranties. All service and repair of CADD-Legacy® 1400 pumps must be performed by Smiths Medical or its authorized agents.
- Review programming screens when complete to make sure desired programming has been entered. Check to make sure unintended changes were not made to the morning dose, continuous rate, or extra dose volume. If unintended changes were made, go to the appropriate screen and program the desired value.
Table of Contents |
Warnings |
|
Cautions |
|
1.0 General Description |
Introduction |
Indications |
Symbols |
Pump Diagram |
Description of the Keys, Display, and Features |
The Main Screen |
Lock Levels |
Security Codes |
Lock Level 0 (LL0) Table |
Lock Levels 1 and 2 (LL1, LL2) Table |
|
2.0 Pump Setup and Programming |
Installing or Replacing the Batteries |
Watching Power Up |
Changing to Lock Level 0 (LL0) |
Programming the Pump: General Instructions |
Delivery Methods |
Programming Screens |
Programming Delivery |
Programming a Morning Dose |
Removing a Medication Cassette Reservoir |
Attaching a Medication Cassette Reservoir |
Priming the Tubing and Connecting to the Patient |
Setting the Lock Level for the Patient |
Programming with Upper Limits, Adjusting Doses in Lock Level 1 |
|
3.0 Operating the Pump |
Starting the Pump |
Stopping the Pump |
Turning the Pump On/Off |
Starting a Morning Dose |
Starting an Extra Dose |
Stopping an Extra Dose or Morning Dose |
Resetting the Reservoir Volume |
|
4.0 Biomed Functions |
Overview: Accessing the Biomed Functions and Programming the Lockouts |
Extra Dose Lockout |
Morning Dose Lockout |
Upstream Occlusion Sensor On/Off |
|
5.0 Reference |
Messages and Alarms, Alphabetical List |
What if the pump is dropped or hit? |
Cleaning the Pump and Accessories |
Exposure to Radiation, Ultrasound, Magnetic Resonance Imaging (MRI), or Use near ECG Equipment |
Continuous Rate Scroll Ranges |
Extra Dose, Morning Dose Scroll Ranges |
Technical Description |
Specifications (Nominal) |
Accuracy Test Results |
Electromagnetic Emissions and Immunity Declarations |
Safety Features and Fault Detection |
Software Safety Features |
Data Handling Software Safety Features |
Annual Functional Inspection |
Collect Separately |
Limited Warranty |
Index |
|
Appendix A – Pump Programming Quick Reference for Healthcare Providers |
1.0 General Description
Introduction
The CADD-Legacy® 1400 pump provides enteral delivery of medication to patients in hospital or outpatient settings. Therapy should always be overseen by a physician or a certified, licensed healthcare professional. As appropriate to the situation, the patient should be instructed in using and troubleshooting the pump.
Indications
The CADD-Legacy® 1400 pump is indicated solely for the enteral delivery of medication contained in a medication cassette reservoir supplied by AbbVie. The medication cassette reservoir attaches to the bottom of the pump.
WARNING: The CADD-Legacy® 1400 pump and medication cassette reservoir are designed for enteral delivery of medication only. They are NOT intended for IV or other parenteral routes of infusion. Use of this product for medications or therapies outside the intended use can result in death or serious patient injury. |
Refer to AbbVie’s full prescribing information for DUOPA (carbidopa and levodopa) enteral suspension for indications and usage, contraindications, warnings, precautions, and adverse reactions.
Symbols
 |
Direct current (power jack) |
 |
Accessory jack |
 |
Caution |
 |
Class II equipment |
 |
Type CF equipment |
 |
Splashproof – water splashed against the pump housing will have no harmful effects (see Cleaning the Pump and Accessories, Section 5, for additional important information). |
 |
Catalog number |
 |
Serial number |
 |
Date of manufacture |
 |
Manufacturer |
 |
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. |
 |
Collect separately |
 |
Temperature limitation |
 |
Humidity limitation |
 |
Atmospheric pressure limitation |
 |
MR Unsafe |
Pump Diagram

Description of the Keys, Display, and Features
AC Indicator Light
The green indicator light is on when you are using the AC adapter to power the pump.
Display
The liquid crystal display (LCD) shows programming information and messages. In this manual, the term “display” is synonymous with display panel or LCD.
Keypad
The keys on the keypad are described below. A key beeps when pressed if it is operable in the current lock level.
 |
used to start and stop pump delivery; silences alarms. |
 |
used to enter (save) a new value in the pump’s memory when programming doses or pump settings or to clear values from record-keeping screens. It is also used to return from the biomed functions to the main screen (see Section 4). |
 |
used to fill the tubing with medication. |
 |
used to confirm and deliver the morning dose (typically used as a daily loading dose) when the pump is running. When the pump is stopped, it is used to view or change the pump’s current lock level. Lock levels are used to limit patient access to certain programming and operating functions. (See Lock Levels, this section.) |
 |
used to move from one programming screen to the next without changing the setting or value displayed; silences alarms. |
 |
used to “scroll up” or increase a value, or scroll through biomed function settings. |
 |
used to “scroll down” or decrease a value, or scroll through biomed function settings. |
 |
used to put the pump into a low power state when not in use or back into full power. |
 |
used by the patient to deliver a programmed amount of medication upon request (extra dose). |
Power Jack
You may plug an AC adapter into the power jack as an alternate source of power. The indicator light on the front of the pump will illuminate when the AC adapter is in use.
Accessory Jack
The accessory jack is used for attaching accessory cables. See the instructions for use supplied with those accessories.
Medication Cassette Reservoir
The medication cassette reservoir is the single-use reservoir designed for use with the CADD-Legacy® 1400 pump. In this manual and on the pump’s display, the word “disposable” refers to the medication cassette reservoir. In AbbVie’s patient instructions for use, medication cassette reservoir is referred to as DUOPA cassette.
Battery Compartment
Two AA batteries fit into the battery compartment. The AA batteries serve as the primary source of power, or as backup power when an AC adapter is in use.
Cassette Latch
The cassette latch attaches the cassette to the pump. The term “cassette” refers to the part of the medication cassette reservoir that attaches to the bottom of the pump. If the cassette becomes unlatched while the pump is running, delivery will stop and an alarm will occur. If the cassette becomes unlatched while the pump is stopped, an alarm will occur.
Other Features Not Shown
Upstream Occlusion Sensor: The pump contains an upstream occlusion sensor. This feature may be turned on or off (see Section 4, Biomed Functions). When the sensor is turned on, and an occlusion in the reservoir is detected, an alarm will sound, delivery will stop, and the display will show Upstream Occlusion.
CAUTION: When the upstream occlusion sensor is turned off, the pump will not detect occlusions in the medication cassette reservoir. Periodically inspect the medication cassette reservoir for decreasing volume. Undetected occlusions could result in under- or non-delivery of medication. |
Downstream Occlusion Sensor: The pump contains a downstream occlusion sensor. When a downstream occlusion (between the pump and the patient) is detected, an alarm will sound, delivery will stop, and the display will show High Pressure.
Reservoir Volume Alarm: The reservoir volume alarm indicates when the volume of medication in the medication cassette reservoir is low or depleted. Each time you change the medication cassette reservoir, you may reset the reservoir volume to the originally programmed value. Then, as medication is delivered, the reservoir volume automatically decreases. When the pump calculates that 5 ml remain in the medication cassette reservoir, beeps sound and ResVol Low appears on the main screen. This alarm recurs at every subsequent decrease of 1 ml until the reservoir volume reaches 0 ml, at which point the pump stops and the reservoir volume empty alarm sounds.
NOTE: The default setting for Reservoir Volume is Not in Use. The reservoir volume alarm is activated only when a value is programmed into the Reservoir Volume screen. Programming a reservoir volume value is not required for general use, but is available at provider discretion.
The Main Screen
The main screen is the starting point for programming or viewing the pump’s settings.
If no keys are pressed for 2 minutes, the display reverts to the main screen. When the two AA batteries are low, LowBat appears on the main screen.

*Does not appear on the main screen if reservoir volume is programmed to Not In Use.

Lock Levels
Lock levels are used to limit patient access to certain programming and operating functions. The table on the next page lists the functions that are accessible in lock level 0 (LL0), lock level 1 (LL1), and lock level 2 (LL2). When a function is accessible, the key associated with the function beeps when pressed. If a function is not accessible, the pump ignores the key press and a beep does not sound. Section 2, Pump Setup and Programming, describes how to change the lock level.
Security Codes
The following security codes are preset by the manufacturer for the clinician’s use:
**Text Omitted**
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment. |
Lock Level 0 (LL0) Table
This table lists the operations that are accessible in lock level 0 (LL0) while the pump is stopped and running. LL0 permits complete access to all programming and operating functions.
|
Stopped |
Running |
Pump Operations and Programming |
LL0 |
Any Lock Level |
Stop/Start the pump |
Yes |
Yes |
Reset reservoir volume |
Yes |
No |
Prime |
Yes |
No |
Change the lock level |
Yes, w/code |
No |
Change morning dose |
No |
Yes |
Start an extra dose |
No |
Yes |
Start a morning dose |
No |
Yes |
Change continuous rate |
Yes |
No |
Change extra dose |
Yes |
No |
Clear given amount |
Yes |
No |
Biomed Functions |
Access to functions |
Yes, w/code |
No |
Extra dose lockout |
Yes, w/code |
No |
Morning dose lockout |
Yes, w/code |
No |
Upstream occlusion Sensor On/Off |
Yes, w/code |
No |
Lock Levels 1 and 2 (LL1, LL2) Table
This table lists the operations that are accessible in lock level 1 (LL1) and lock level 2 (LL2) while the pump is stopped and running. LL1 permits limited control of pump programming and operations. LL2 permits only minimal control of pump operations.
Pump Operations and Programming |
Stopped |
Running |
|
LL1 |
LL2 |
Any Lock Level |
Stop/Start the pump |
Yes |
Yes |
Yes |
Reset reservoir volume |
Yes |
Yes |
No |
Prime |
Yes |
No |
No |
Change the lock level |
Yes, w/code |
Yes, w/code |
No |
Change morning dose |
No |
No |
Yes* |
Start an extra dose |
No |
No |
Yes |
Start a morning dose |
No |
No |
Yes |
Change continuous rate |
Yes* |
No |
No |
Change extra dose |
Yes* |
No |
No |
Clear given amount |
Yes |
No |
No |
*When in LL1, you can program up to the LL0 value. No programming is allowed in LL2.
2.0 Pump Setup and Programming
Installing or Replacing the Batteries
Use new, AA alkaline batteries such as DURACELL® or EVEREADY® ENERGIZER® batteries to power the pump. The pump retains all programmed values while the batteries are removed.
Dispose of used batteries in an environmentally safe manner, and according to any regulations which may apply.
WARNING: |
- Always have new batteries available for replacement. If power is lost, non-delivery of medication will occur, which could compromise patient treatment.
|
- If the pump is dropped or hit, the battery door or tabs may break. Do not use the pump if the battery door or tabs are damaged because the batteries will not be properly secured; this may result in loss of power and non-delivery of medication, which could compromise patient treatment.
|
 |
CAUTION: Do not use rechargeable NiCd or nickel metal hydride (NiMH) batteries. Do not use carbon zinc (“heavy duty”) batteries. They do not provide sufficient power for the pump to operate properly. |
In order to install or replace the batteries, be sure the pump is Stopped. Then, follow these steps:
1. Push down and hold the arrow button while sliding the door off.

2. Remove the used batteries. Pulling on the end of the battery strap will make battery removal easier.
3. Install the new batteries in the compartment, making sure the battery strap is positioned correctly under the batteries.

NOTE:
- Be sure to match the polarity markings of the new batteries (+ and –) with those labeled in the battery compartment. If you put the batteries in backwards, the display will remain blank, and you will not hear a beep.
- Use two new, AA alkaline batteries to power the pump. You may use any alkaline batteries, including DURACELL® Alkaline and EVEREADY® ENERGIZER® Alkaline, for example.
4. Place the battery door over the battery compartment and slide the door closed.

5. Ensure that the door is latched by trying to remove the door without pressing the arrow button.
NOTE: The power-up sequence will start, the pump will go through an electronic self-test, and the pump will beep 6 times at the end of the power-up sequence. All of the display indicators, the software revision, and each parameter will appear briefly.
WARNING: If a gap is present anywhere between the battery door and the pump housing, the door is not properly latched. If the battery door becomes detached or loose, the batteries will not be properly secured; this could result in loss of power and non-delivery of medication, which could compromise patient treatment. |

6. Resume operation of the current program by pressing and holding

to start the pump or proceed to program the pump.
NOTE:
- The life of the batteries is dependent on the amount of medication delivered, delivery rate, battery age, and the temperature.
- At the rate of 100 ml per day, alkaline batteries will usually last about 7 days.
- The power of the batteries will be quickly depleted at temperatures below 10°C (50°F).
CAUTION: Do not store the pump for prolonged periods of time with the batteries installed. Battery leakage could damage the pump. |
Watching Power Up
When you install the batteries, the pump will start its power up sequence during which it performs self-tests and displays programmed values. Watch for the following:
- Pump model number and last error code (“LEC”) if any, will appear. (If an error code appears, the pump should be removed from use and returned for service.)
- The software revision will appear.
- The display will turn on, showing a series of blocks. Look for any blank areas, which would indicate a faulty display.
- The display will turn off briefly.
- The pump’s program screens will appear, followed by the current lock level setting. The pump will beep after each screen. If messages appear, see Messages and Alarms Table, Section 5 for further explanation and instruction.
- When power up is complete, 6 beeps will sound, and the pump will be stopped on the main screen.
NOTE: To move quickly through the power-up screens, press

repeatedly. To skip the automatic review entirely, press

If you attempt to skip screens before the pump is powered up, it will not respond.
Changing to Lock Level 0 (LL0)
Before programming the pump, make sure the pump is set to LL0. LL0 allows the clinician to access all programming and operating functions.
- Make sure the pump is stopped. Press

The current lock level will appear. (If the lock level is already LL0, press

to exit.)
**Text Omitted**
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment. |
Programming the Pump: General Instructions
WARNING: System delivery inaccuracies beyond the stated accuracy may occur as a result of back pressure or fluid resistance, which depends upon temperature, medication viscosity, catheter size, extension set tubing, and orientation of the pump system. |
The procedure for changing a programmed setting is similar for most programming screens.
- Make sure the pump is stopped and in lock level 0.
- To begin programming, start at the main screen and press

- To change a setting, press

or

until the desired setting appears. (Press and hold these keys to change values with increasing speed.)

within 25 seconds to confirm a change or the screen will revert to the previous setting.

is pressed, Value not saved will appear. Press

to return to the screen being programmed, scroll to the desired value, and press


to advance to the next screen.
- To leave a setting unchanged, press

to go to the next screen.
Delivery Methods
WARNING: Programming the pump at a delivery rate other than what is prescribed will cause over- or under-delivery of medication, which could compromise patient treatment. Please refer to the prescribing information for DUOPA for dosage and administration information. |
The CADD-Legacy® 1400 pump offers 3 methods of delivery:
- Continuous rate
- Extra dose
- Morning dose
The following graph illustrates the combined delivery methods. The continuous rate, extra dose, and morning dose are programmed as described in this section. Ranges and programming increments are listed in the Specifications in Section 5.

Programming Screens
These are the programming screens for the CADD-Legacy® 1400 pump. Descriptions of the screens follow.
Reservoir Volume (ml) |
 |
Continuous Rate (ml/hr) |
 |
Extra Dose (ml) |
 |
Given (ml) |
 |
Reservoir Volume
NOTE: The default setting for Reservoir Volume is Not in Use. Programming a reservoir volume value is not required for general use, but is available at provider discretion.
If you wish to use the reservoir volume feature, enter the volume of medication contained in the filled medication cassette reservoir. The reservoir volume value decreases as the pump delivers medication or as you prime the tubing. When you change the medication cassette reservoir, reset the reservoir volume value on this screen. If you do not wish to use the reservoir volume feature, scroll down to Not In Use (located before 1 and after 9999 in the range of values).
The reservoir volume value could be set higher than the capacity of the medication cassette reservoir. Be sure to program the reservoir volume to reflect the actual volume in the reservoir.
Continuous Rate
Enter the continuous rate of medication delivery in ml/hr. The maximum rate is 20 ml/hr. If the prescription does not call for a continuous rate, enter zero.
NOTE: If you intend to run the pump in lock level 1 so the continuous rate can be varied, you should enter the maximum allowable rate while programming in lock level 0. After programming, you may then change to lock level 1 and decrease the rate to its starting value. See Programming with Upper Limits, Adjusting Doses in LL1 at the end of Section 2.
Extra Dose
Enter the amount of medication to be delivered when the patient presses

If the prescription does not call for an extra dose, enter zero.
NOTE: If you intend to run the pump in lock level 1 so the extra dose can be varied, you should enter the maximum allowable dose while programming in lock level 0. After programming, you may then change to lock level 1 and decrease the dose to its starting value. See Programming with Upper Limits, Adjusting Doses in LL1 at the end of Section 2.
Given
This screen shows the total amount of medication delivered since the last time this value was cleared. The amount shown is rounded to the nearest 0.05 ml. If this value reaches 99999.95, it automatically returns to 0 and continues counting. When using the pump’s

key, the amount of medication used is not included in the Given amount.
Other Programming Information
The morning dose should be programmed separately following programming of the above. Information on programming the morning dose can be found later in this section.
Programming Delivery
WARNING: Programming the pump at a delivery rate other than what is prescribed will cause over- or under-delivery of medication, which could compromise patient treatment. Please refer to the prescribing information for DUOPA for dosage and administration information. |
To program the pump, enter the prescribed values.
1. Begin at the main screen.
- Make sure the pump is in LL0.
- Make sure STOPPED appears on the main screen.
- Press

to begin.
2. Enter the reservoir volume (optional – not required for general use).
NOTE: The default setting for Reservoir Volume is Not in Use. Programming a reservoir volume value is not required for general use, but is available at provider discretion.

or

to select the volume in the filled medication cassette reservoir. (If you do not wish to use the reservoir volume feature, scroll down to Not In Use located before 1 or after 9999.)


3. Enter the continuous rate.

or

to select the desired continuous rate.


4. Enter the extra dose amount.

or

to select the desired extra dose amount.

NOTE: If required, program the extra dose lockout time, as instructed in Section 4, Biomed Functions.
5. Clear the amount given.

if you wish to clear the amount given.

6. Review the program.
Press

repeatedly to review the programming screens. If you need to reprogram a setting, press

until the appropriate screen appears and change the setting as described in this section.
Programming a Morning Dose
To program a morning dose the pump must be running and a medication cassette reservoir must be attached.
To program a morning dose
1. Make sure the pump is running and in LL0 or LL1. Start the pump, if necessary.
NOTE: In LL0, programming in the full range is possible. In LL1, you can program up to the LL0 value.
2. Press

The current morning dose value will appear.

3. Press

or

to select the desired morning dose value.
4. Press

to enter the value into the pump’s memory. (If desired, press

again to begin delivering the morning dose.)
Once entered, the morning dose amount is retained in the pump’s memory. The patient can then press

twice to display and deliver the morning dose.
NOTE: If required, program the morning dose lockout time, as instructed in Section 4, Biomed Functions.
Removing a Medication Cassette Reservoir
WARNING: Clamp the fluid path tubing and/or disconnect the tubing from the enteral access device before removing the medication cassette reservoir from the pump to prevent uncontrolled delivery of medication, which could compromise patient treatment. |
To remove the medication cassette reservoir from the pump
1. Stop the pump.
2. Close the tubing clamp. If necessary, disconnect the tubing from the enteral access device.
3. Insert a coin into the slot in the cassette latch and turn it clockwise. The latch will pop out when you unlatch the cassette (the part of the medication cassette reservoir that attaches to the bottom of the pump).
4. A continuous alarm will sound and the pump will display No Disposable, Clamp Tubing (the pump is not sensing proper cassette attachment). The alarm may be silenced by pressing

or

5. Remove the cassette hooks from the pump hinge pins.
Attaching a Medication Cassette Reservoir
Obtain a new, filled medication cassette reservoir.
WARNING: |
- Use only approved DUOPA medication cassette reservoirs to maintain pump accuracy and assure proper pump operations.
- Use only extension sets approved for use with DUOPA, paying particular attention to all warnings and cautions associated with their use.
|
After attaching the medication cassette reservoir, proceed to the reservoir volume screen to reset the value for the volume, and then prime the tubing.
To attach the medication cassette reservoir to the pump |
1. Clamp the tubing.
2. Insert the cassette hooks into the hinge pins on the pump.
3. Place the pump upright on a firm, flat surface. Press down so the cassette fits tightly against the pump. |
 |
4. Insert a coin into the slot in the cassette latch, push in, and turn counterclockwise until the line on the latch lines up with the arrow on the side of the pump and you feel the latch click into place. |
 |
WARNING: Attach the cassette properly. The cassette is the part of the medication cassette reservoir that attaches to the pump. A detached or improperly attached cassette could result in unintended delivery of medication, which could compromise patient treatment. |
5. Gently twist, push, and pull on the medication cassette reservoir to make sure it is firmly attached. If the cassette is not secure, repeat the procedure from step 1. |
 |
Priming the Tubing and Connecting to the Patient
WARNING: Do not prime the fluid path with the tubing connected to a patient as this could result in over-delivery of medication, which could compromise patient treatment. |
The pump must be stopped and in LL0 or LL1 in order to prime the fluid path. If the pump is in LL2, you cannot prime the fluid path.
NOTE: If you are not changing the medication cassette reservoir but wish to prime the fluid path, you may follow the same procedure.
1. Make sure the tubing is disconnected from the patient and the tubing clamp is open.
2. Press and hold

You will hear a single beep, and the word Prime and 3 sets of dashes, each accompanied by a beep, will appear on the display.
3. After Prime and 3 sets of dashes appear, release

4. Press and hold

again to fill the fluid path. The screen displays Priming . . . and you will hear a short beep each time the pump goes through a delivery cycle.
NOTE: Medication delivered during priming is subtracted from the reservoir volume, but is not added to the given screen since this amount is not delivered to the patient.
5. If the tubing is not yet fully primed, press and hold

again. If the tubing is primed, press

to return to the main screen.
NOTE: Each time you press and hold

you pump a maximum of 1 ml of medication into the tubing. The pumping action will stop automatically when 1 ml has been delivered. Release

if you finish priming the fluid path sooner. If the fluid path is not fully primed, repeat the above priming procedure.
6. Connect the tubing to the patient’s enteral access device.
7. Set the lock level for the patient (see Setting the Lock Level for the Patient in this section).
Setting the Lock Level for the Patient
The lock level must be changed to LL1 or LL2 to prevent the patient from having complete access to all programming and operating functions.
NOTE: You may change the lock level at any time by stopping the pump and following the procedure below.
To change the lock level
**Text Omitted**
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment. |
Programming with Upper Limits, Adjusting Doses in Lock Level 1
If a prescription allows for the continuous rate, extra dose, or morning dose to be adjusted during the course of therapy, you may wish to operate the pump in LL1. Then, when necessary, you can adjust values up to the maximum value that was programmed in LL0.
Programming the pump to use this feature
- During initial programming in LL0, enter the upper limit values for the continuous rate, extra dose and/or morning dose. (These will be the maximum values when the pump is in LL1.)
- After you are finished programming, change the lock level to LL1.
- Decrease the continuous rate, extra dose and/or morning dose to its starting value, then press

Adjusting the rate or dose while the pump is in use
- If it becomes necessary to increase or decrease the continuous rate, and/or extra dose during the course of therapy, stop the pump but remain in LL1.
- To increase or decrease the morning dose, the pump must be in LL1, but it must remain running.
1. To change the continuous rate or extra dose, press

until the continuous rate or extra dose screen appears.
To change the morning dose, press

2. Press

or

to select the desired value, then press

NOTE: You will not be able to adjust the continuous rate, extra dose, or morning dose beyond the value originally programmed in LL0.
3. Restart the pump if appropriate.
3.0 Operating the Pump
Starting the Pump
When you start the pump, programmed values will be automatically reviewed. Then medication delivery will begin as programmed, and RUN will appear on the main screen. If the pump will not start, a message will appear on the display. Refer to the Messages and Alarms Table in Section 5.
CAUTION: Prior to starting medication delivery, inspect the fluid path for kinks, a closed clamp, or other obstruction. An undetected occlusion may result in under- or non-delivery and/or nuisance alarms. |
To start the pump
1. Press and hold

Starting and 3 sets of dashes appear on the display; then the dashes disappear one-by-one, each accompanied by a single beep.
2. Release

after the last set of dashes disappears, and the pump beeps. All of the programming screens appear for your review one after the other.
Stopping the Pump
Stopping the pump stops delivery. When the pump is stopped, STOPPED will appear on the main screen, and you will hear 3 beeps every 5 minutes.
To stop the pump
1. Press and hold

Stopping and 3 sets of dashes appear one-by-one on the pump’s display, each accompanied by a single beep.
2. Release

after the third set of dashes appears and the pump beeps.
Turning the Pump On/Off
When the pump is stopped, you may put the pump into a low power state by turning it off. The pump may be turned off when it is disconnected from the patient and it is going to be stored for short periods of time.
CAUTION: Do not store the pump for prolonged periods of time with the batteries installed. Battery leakage could damage the pump. |
To turn the pump off
1. Press and hold

Turning off and 3 sets of dots will appear one-by-one on the pump’s display, each accompanied by a single beep.
2. Release

The display turns off.
To turn the pump on
1. Press and hold

until the display turns on.
2. Release

The pump will power up and automatically review all screens.
Starting a Morning Dose
A morning dose may be delivered in any lock level while the pump is running. It allows you to deliver a specified amount of medication, as a loading dose for example.
If the patient attempts to deliver a morning dose during the lockout time, the pump will not deliver the dose. The lockout time is determined by the value entered in Morning Dose Lockout, in biomed functions. The extra dose lockout setting has no effect on morning dose frequency. A morning dose may be stopped in progress.
NOTE: A morning dose cannot be started while an extra dose or another morning dose is in progress.
To start a morning dose
To program a morning dose, the pump must be running and a medication cassette reservoir must be attached.
NOTE: The

key must be pressed twice for morning dose delivery to start.
- Make sure the pump is running (in any lock level). Start the pump if necessary.
- Press

The current morning dose value (or the default value of 0 ml) will appear UNLESS the morning dose is currently locked out (in which case the screen will not appear). If the desired morning dose amount appears in the display, press

again to begin delivery.
NOTE: If the desired morning dose amount does not appear in the display, program the desired morning dose amount as instructed in Section 2, Pump Setup and Programming.
The screen will show the value decreasing as the morning dose is delivered.
Starting an Extra Dose
If an extra dose has been programmed, the patient may start an extra dose while the pump is running. The amount delivered is added to the amount provided by the continuous rate.
If the patient attempts to deliver an extra dose during the lockout time, the pump will not deliver the dose. The lockout time is determined by the value entered in Extra Dose Lockout, in biomed functions.
NOTE: An extra dose cannot be started while another extra dose or a morning dose is in progress.
To start an extra dose
- Make sure the pump is running (in any lock level). Start the pump if necessary.
- Press

Two beeps will sound and the pump will begin delivering the extra dose.
As the extra dose is delivered, the main screen will show DOSE in place of RUN.
Stopping an Extra Dose or Morning Dose
An extra dose or morning dose can be stopped in progress. The pump may be in any lock level.
To stop an extra dose or morning dose in progress

to stop the pump. All delivery is stopped, including the continuous rate.
Resetting the Reservoir Volume
NOTE: The default setting for Reservoir Volume is Not in Use. Programming a reservoir volume value is not required for general use, but is available at provider discretion.
To reset the reservoir volume to the value programmed in LL0, the pump may be in any lock level.
1. Stop the pump.
2. Press

to display the reservoir volume screen.
3. Press

to reset the volume to the programmed value.
4.0 Biomed Functions
Overview: Accessing the Biomed Functions and Programming the Lockouts
The biomed functions are pump configurations that are less frequently changed. The biomed functions are accessible only when the pump is stopped and in lock level 0 (LL0).
To access the Biomed Functions
**Text Omitted**
WARNING: Do not disclose to the patient the pump’s security codes or any other information that would allow the patient complete access to all programming and operating functions. Improper programming could compromise patient treatment. |
4. Press

to select the setting you wish to view or change, then follow the instructions in this section for the appropriate screen.
NOTE: To leave a biomed function setting unchanged, press

5. To exit the biomed functions, press

until you get to the screen that reads, NEXT for Biomed, ENTER for Main.
6. Press

to return to the main screen.
Extra Dose Lockout
The extra dose lockout time determines how often a patient can receive an extra dose of medication. The lockout time is the minimum amount of time which must elapse between the start of one dose and the start of the next.
To program an extra dose lockout time |
1. With the pump stopped and in LL0, access biomed functions. (Refer to the beginning of the Biomed Functions section for instructions on how to access biomed functions.) |
2. Press
 until Extra Dose appears.
3. Press
 or
 to scroll to the desired lockout time. The default lockout time is 1 hour.
4. Press
 to enter the value. |
 |
CAUTION: When you enter a new value, any lockout time already in effect will be cleared. An extra dose could be requested and delivered immediately upon starting the pump, which may result in over-delivery of medication. |
Morning Dose Lockout
The morning dose lockout time determines how often a patient can receive a morning dose. The lockout time is the minimum amount of time which must elapse between the start of one dose and the start of the next.
To program a morning dose lockout time |
1. With the pump stopped and in LL0, access biomed functions. (Refer to the beginning of the Biomed Functions section for instructions on how to access biomed functions.) |
2. Press
 until Morning Dose appears.
3. Press
 or
 to scroll to the desired lockout time.
4. Press
 to enter the value. |
 |
CAUTION: When you enter a new value, any lockout time already in effect will be cleared. A morning dose could be requested and delivered immediately upon starting the pump, which may result in over-delivery of medication. |
Upstream Occlusion Sensor On/Off
The upstream occlusion sensor can be set to On or Off. If this screen is set to On, and an occlusion in the medication cassette reservoir is detected, an alarm will sound, delivery will stop, and the display will show Upstream Occlusion.
CAUTION: When the upstream occlusion sensor is turned off, the pump will not detect occlusions in the medication cassette reservoir. Periodically inspect the medication cassette reservoir for decreasing volume. Undetected occlusions could result in under- or non-delivery of medication. |
1. With the pump stopped and in LL0, access biomed functions. (Refer to the beginning of the Biomed Functions section for instructions on how to access biomed functions.) |
2. Press
 until Upstream Sensor appears.
3. Press
 or
 to select Off or On.
4. Press
 to enter the change. |
 |
5.0 Reference
Messages and Alarms, Alphabetical List
Messages and Alarms |
Description / Corrective Action |
[No message] Two-tone alarm |
With no AC adapter attached, the batteries have been removed while the pump is running. The pump is now stopped and unpowered. Install batteries to silence the alarm. OR Batteries were removed within approximately 15 seconds after stopping the pump. Install new batteries to silence the alarm, if desired. Otherwise, the alarm will stop within a short period of time. |
[Screen displays current pump status] Two-beeps (long-short) |
The medication cassette reservoir is not aligned with the pump or medication is not flowing from the medication bag inside the reservoir to the pumping mechanism. Press
 or
 to silence the alarm. The pump continues to run. Make sure the medication cassette reservoir is properly aligned with the pump and medication is flowing from the medication bag to the pumping mechanism. Very cold or extremely thick medication may cause this alarm as well. Allow the medication cassette reservoir to thaw to room temperature before attaching to the pump. |
Battery Depleted Two-tone alarm |
The battery power is too low to operate the pump. The pump is now stopped.
- Change the batteries immediately.
- Press and hold
 to restart the pump. |
Battery Removed Pump won’t run Two-tone alarm |
With the AC adapter attached, the AA batteries have been removed while the pump is running, or you have tried to start the pump with depleted batteries. The pump is now stopped. Press
 or
 to silence the alarm. Reinstall batteries or install new batteries. Press and hold
 to restart the pump. |
Error Two-tone alarm |
An error has occurred. Remove the pump from service and contact Smiths Medical to return the pump for service. |
High Pressure Two-tone alarm |
The pump has detected high pressure, which may be resulting from a downstream blockage, kink in the fluid path, or a closed clamp. Remove the occlusion to resume operation. Or, press
 or
 to stop the pump and silence the alarm for 2 minutes, then remove the occlusion and restart the pump. |
Key pressed, Please release Two-tone alarm |
If a key is being pressed, stop pressing it. If the alarm persists, close the tubing clamp and remove the pump from use. Contact Smiths Medical to return the pump for service. |
LowBat Three two-tone beeps every 5 minutes |
The batteries are low, but the pump is still operable.
- Change the batteries soon.
|
Motor Locked, remove all power Two-tone alarm |
Batteries are depleted and the pump was powered up with the AC adapter. Install new AA batteries, reconnect the AC adapter, and restart the pump. |
No Disposable, Clamp Tubing Two-tone alarm |
The disposable (medication cassette reservoir) was removed. The pump is not sensing proper cassette attachment. Clamp the tubing and disconnect from the enteral access device. A medication cassette reservoir must be properly attached in order for the pump to run. Press
 or
 to silence the alarm. |
No Disposable, Pump won’t run Two-tone alarm |
You have tried to start the pump without a disposable (medication cassette reservoir) attached. A medication cassette reservoir must be properly attached in order for the pump to run. Press
 or
 to silence the alarm. |
Power lost while pump was on Two-tone alarm |
The pump was running when power was removed. Stop the pump before changing the batteries or removing the power source. Press
 or
 to silence the alarm. |
Programming Incomplete Two-tone alarm when starting the pump |
A rate or dose must be programmed to start the pump. Press
 or
 to silence the alarm. |
Reservoir Volume Empty Two-tone alarm |
The reservoir volume has reached 0.0 ml. Press
 or
 to silence the alarm. Then install a new medication cassette reservoir if appropriate and reset the reservoir volume. |
RUN ResVol Low Three single beeps |
The reservoir volume is low. Change the medication cassette reservoir soon. See Reservoir Volume Alarm in Section 1 for further details. |
Service Due Two-tone alarm |
The pump is functional, but is due for service based on clock battery age or total motor revolutions. This screen appears in LL0 for 60 days and then in all lock levels until returned for service. |
Upstream Occlusion Two-tone alarm |
If the upstream occlusion sensor is set to On and an occlusion in the medication cassette reservoir is detected, the upstream occlusion alarm will sound. Press
 or
 to stop the pump and silence the alarm for 2 minutes, then remove the occlusion and restart the pump. You may have to detach the cassette from the pump, then reattach it. |
Value not saved |
A value was not saved by pressing
 Press
 to resume programming. Verify all programming screens before moving to the next screen or starting the pump. |
What if the pump is dropped or hit? |
Immediately do the following: |
- Check the cassette latch on the side of the pump and make sure the line on the latch lines up with the arrow on the side of the pump.
|
 |
- Gently twist, push, and pull on the medication cassette reservoir to make sure it is still firmly attached.
|
 |
- Check the battery door to make sure it is still firmly attached.
|
If the medication cassette reservoir or the battery door is loose or damaged, do not use the pump. Immediately stop the pump, close the tubing clamp, and contact Smiths Medical.
WARNING: If the pump is dropped or hit, inspect the pump for damage. Do not use a pump that is damaged or is not functioning properly, as this could compromise patient treatment. |
- If the pump is accidentally dropped in water, retrieve it quickly, dry it off with a towel, and contact Smiths Medical.
Cleaning the Pump and Accessories
CAUTION: |
- Do not immerse the pump in cleaning fluid or water. Do not allow solution to soak into the pump, accumulate on the keypad, or enter the battery compartment. Moisture buildup inside the pump may damage the pump.
- Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners, as damage to the pump may occur.
|
NOTE: Refer to the Instructions for Use for each accessory before proceeding with cleaning.
The following solutions may be used to clean the pump and accessories, unless otherwise specified:
- Soap solution
- Benzalkonium chloride concentrate (0.13%)
- Glutaral concentrate, USP (2%)
- 10% solution of household bleach (one part household bleach to 9 parts water)
- Alcohol, USP (93%)
- Isopropyl Alcohol, USP (99%)
- Chlorhexidine gluconate (4%)
- PDI Super Sani-Cloth®
- MadaCide, Mada Medical
- Dampen a soft, lint-free cloth with cleaning solution. Apply the solution to the exterior surface of the pump (per manufacturer’s instructions). Do not allow the solution to soak into the pump or accessory.
- Wipe the entire surface dry with another soft, lint-free cloth. Allow the pump to dry completely before use.
Cleaning the Battery Contacts
Routinely clean the battery contacts, possibly as part of the preventative maintenance cycle, to remove buildup of foreign material on the contacts.
Use the following to clean the battery contacts:
- Cotton swab wetted with Isopropyl Alcohol (70% minimum)
NOTE: Do not use an alcohol formulation that contains components other than alcohol and water.
OR
- Pre-moistened alcohol swab
- Using a swab wetted with alcohol, rub the entire battery contact for a minimum of 10 back and forth cycles (20 total wipes over the contact).
- Using a clean surface of the swab, repeat process for second battery contact.
- Using a clean swab wetted with alcohol, rub each battery contact again, a minimum of 4 back and forth cycles (8 total wipes over the contact).
- Allow the contacts to dry completely before use.
Exposure to Radiation, Ultrasound, Magnetic Resonance Imaging (MRI), or Use near ECG Equipment
CAUTION: |
- Do not expose the pump to therapeutic levels of ionizing radiation as permanent damage to the pump’s electronic circuitry may occur. The best procedure to follow is to remove the pump from the patient during therapeutic radiation sessions. If the pump must remain in the vicinity during a therapy session, it should be shielded, and its ability to function properly should be confirmed following treatment.
- Do not expose the pump directly to ultrasound, as permanent damage to the electronic circuitry may occur.
- Do not use the pump in the vicinity of magnetic resonance imaging (MRI) equipment as magnetic fields may adversely affect the operation of the pump. Remove the pump from the patient during MRI procedures and keep it at a safe distance from magnetic energy.
- This pump may interfere with ECG equipment. Monitor ECG equipment carefully when using this pump.
|
Continuous Rate Scroll Ranges
Starting Value |
Increment |
Maximum |
0.0 ml/hr |
0.1 ml/hr |
20.0 ml/hr |
Extra Dose, Morning Dose Scroll Ranges
Extra Dose |
increment |
max. |
0.1 ml |
9.9 ml |
Morning Dose |
increment |
max. |
0.1 ml |
20 ml |
Technical Description
Standards used in Development of the Pump |
The following standards were used in whole or part in the development of the pump: |
IEC 60601-1 (1988), (2nd Edition, 1988) Medical Electrical Equipment, Part 1: General Requirements for Safety. Amendment 1 (1991), Amendment 2 (1995). |
EN 60601-1 (1990), Medical Electrical Equipment, Part 1: General Requirements for Safety. Amendment A1 (1993), Amendment A13 (1996), Amendment A2 (1995). |
IEC 60601-1-2 (2007), Medical Electrical Equipment, Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests. |
EN 60601-1-2 (2007), Medical Electrical Equipment, Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests. |
The following are reference test methods applied to IEC/EN 60601-1-2: |
|
IEC/EN 61000-4-2, Electromagnetic Compatibility (EMC), Part 4: Testing and measurement techniques. Section 2: Electrostatic Discharge immunity test. Basic EMC Publication. |
|
IEC/EN 61000-4-3, Electromagnetic Compatibility (EMC), Part 4: Testing and measurement techniques. Section 3: Radiated, radio frequency, electromagnetic fields immunity test. Basic EMC Publication. |
|
IEC/EN 61000-4-4, Electromagnetic Compatibility (EMC), Part 4: Testing and measurement techniques. Section 4: Electrical fast transients/bursts immunity test. Basic EMC Publication. |
|
IEC/EN 61000-4-5, Electromagnetic Compatibility (EMC), Part 4: Testing and measurement techniques. Section 5: Surge immunity test. Basic EMC Publication. |
|
IEC/EN 61000-4-8, Electromagnetic Compatibility (EMC), Part 4: Testing and measurement techniques. Section 8: Power frequency magnetic field immunity test. |
|
IEC/EN 61000-4-11, Electromagnetic compatibility (EMC) - Part 4: Testing and measuring techniques - Section 11: Voltage dips, short interruptions and voltage variations immunity tests. |
IEC 60601-2-24 (1998), Medical Electrical Equipment, Part 2-24: Particular Requirements for Safety of Infusion Pumps and Controllers. |
EN 60601-2-24 (1998), Medical Electrical Equipment, Part 2-24: Particular Requirements for Safety of Infusion Pumps and Controllers. |
IEC 60601-1-4 (2000), Medical Electrical Equipment, Part 1: General Requirements for Safety – Collateral Standard: Programmable electrical medical systems. |
EN 60601-1-4 (1996), Medical Electrical Equipment, Part 1-4: General Requirements for Safety – Collateral Standard: Programmable electrical medical systems. Amendment A1: 1999. |
EN 980 (2008), Graphical symbols for use in the labeling of medical devices. |
FCC Part 15 Subpart B, Radiofrequency Devices, Unintentional Radiators. |
RTCA/DO -160C, Radiated Emissions Only, Category A & Z Limit. |
EN 55011 (2007), Limits and Methods of Measurement of Radio Interference Characteristics of Industrial, Scientific and Medical (ISM) Equipment. Amendment A2: 2007. (Equivalent to CISPR 11: 2003 + Amendment A2: 2006). |
CISPR11 (2009), Limits and methods of measurement of electromagnetic disturbance characteristics of industrial, scientific and medical (ISM) radio frequency equipment. Amendment 1 (1996) Amendment 2 (1996). |
|
For CISPR11 tests, the pump was fitted with an administration set with its inlet connected to a 250 ml bag and its outlet routed back to the bag forming a closed loop system. A total of 6 feet of tubing was used to form the closed loop. |
Specifications (Nominal)
General Pump Specifications |
Resolution |
medication cassette reservoir: 0.05 ml per pump activation nominal |
Size |
4.1 cm × 9.5 cm × 11.2 cm [1.6 in. × 3.8 in. × 4.4 in.] excluding medication cassette reservoir |
Weight |
392 g [13.8 oz.] including 2 AA batteries, empty 100 ml medication cassette reservoir, excluding other accessories |
Classification (IEC 60601-1) |
CF
 Class II
 |
Moisture protection |
Splashproof (IPX4) |
Pump alarms |
Low battery power; depleted battery power; battery dislodged; pump stopped; pump fault; low reservoir volume; high pressure; disposable not attached when run attempted; motor locked; upstream occlusion; reservoir volume empty; program incomplete; remote dose cord removed; key stuck; disposable detached. |
Maximum infusion pressure |
40 psi [2.76 bar] |
Maximum time to occlusion alarm (actual test data) |
DUOPA (carbidopa and levodopa) enteral suspension: 10 hours 30 minutes |
Bolus volume at occlusion alarm pressure |
0.05 ml resolution medication cassette reservoir: <1.1 ml |
Power sources |
AC adapter; Two AA alkaline batteries |
Battery life |
The expected life of 2 AA batteries is 100 hours at 100 ml/16 hours, or approximately 14 days at 10 ml/day (nominal). This estimate is based on laboratory tests conducted at room temperature using two new batteries. Actual battery life will vary depending on the brand of battery, battery shelf life, temperature conditions, and delivery rate. It is recommended that two new AA batteries be kept available for replacement if necessary.
An internal battery powers the clock. When it is depleted, it cannot reliably maintain the clock time. This battery must be replaced by the manufacturer. The internal battery has an expected life of 5 years. |
System operating temperature |
2°C to 40°C (36°F to 104°F) |
System storage and transportation temperature |
-20°C to 60°C (-4°F to 140°F) |
When shipping, use pump case. |
System delivery accuracy (DUOPA [carbidopa and levodopa] enteral suspension): |
Therapy1 |
±10% |
Continuous rate2 |
±10% @ flow rates 0.4 ml/hr to 20 ml/hr |
Extra dose2 |
±10% |
Morning dose2,3 |
±10% |
Accessories: 100 ml medication cassette reservoir with 36” 0.10” I.D. tube |
1Therapy delivery accuracy: program to deliver a 10 ml morning dose with a 5 ml/hr continuous rate and three 3 ml extra doses, with a total of 100 ml of medication delivered. Pumps were evaluated in latch down, upside down, and horizontal position. The medication cassette with DUOPA (carbidopa and levodopa) enteral suspension is initially cold (approx. 2°C). Remove cassette from refrigerator, wait 20 minutes, then conduct the test at nominal temperature.2 2At nominal temperature (23°C ± 5°C) 3An additional -4% change may be seen when cassette starting temperature is 2°C and placed at room temperature (23°C ± 5°C) for 20 minutes prior to use. |
WARNING: System delivery inaccuracies beyond the stated accuracy may occur as a result of back pressure or fluid resistance, which depends upon temperature, medication viscosity, catheter size, extension set tubing, flow rate, and orientation of the pump system. |
System definition |
System is defined as a CADD-Legacy® 1400 pump with attached medication cassette reservoir supplied by AbbVie |
High pressure alarm |
26 ± 14 psi [1.79 ± 0.97 bar] |
Maximum volume infused under single fault condition |
medication cassette reservoir: 0.2 ml. |
Delivery rate during priming |
Approx. 125 ml/hr |
Delivery Specifications |
Reservoir volume |
1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: Not In Use |
Continuous rate |
0 to 20 ml/hr in 0.1 ml/hr increments Default: 0.1 ml/hr |
Extra dose |
0 to 9.9 ml in 0.1 ml increments Delivery rate (continuous rate + extra dose): 40 ml/hr Default: 0 ml |
Given |
0 to 99999.95 in 0.05 unit increments |
Morning dose |
0 to 20.00 ml in 0.1 ml increments Delivery rate (continuous rate + morning dose): 40 ml/hr Default: 0 ml |
Biomed Functions |
Extra dose lockout |
15 minutes to 24 hours in 15 minute increments Default: 1 hour |
Morning dose lockout |
1 to 24 hours in 1 hour increments Default: 20 hours |
Upstream sensor |
Off On Default: Off |
Accuracy Test Results
The following graphs are designed to show flow accuracy of the infusion system plotted against given time periods. The medication cassette reservoir used for flow accuracy tests was supplied by AbbVie.
Flow rate immediately following startup |
Time Interval: |
0.5 min |
Total Time: |
120 min |
Programmed Rate: |
5 ml/hr |
Reservoir Used: |
100 ml medication cassette reservoir |

Short term flow rate error |
Programmed Rate: |
5 ml/hr |
Average Flow Rate: |
4.76 ml/hr |
Mean Flow Error: |
–4.65% |
Reservoir Used: |
100 ml medication cassette reservoir |

Electromagnetic Emissions and Immunity Declarations
Electromagnetic emissions declaration |
The Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Pump should assure that it is used in such an environment. |
Emissions test |
Compliance |
Electromagnetic environment – guidance |
RF emissions, CISPR 11 |
Group 1 |
The Pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
RF emissions, CISPR 11 |
Class B |
The Pump is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
Harmonic emissions IEC 61000-3-2 |
Not applicable |
|
Voltage fluctuations/ flicker emissions IEC 61000-3-3 |
Not applicable |
|
Compliance using:
- 115VAC/60HZ to 8VDC Power Adapter (US) with a cord length of 274 ± 10 cm (108 ± 4 in).
- Remote Dose Cord with a length of 152 ± 5 cm (60 ± 2 in).
WARNING: |
- The Pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the user should verify normal operation of the pump in the configuration in which it is to be used.
- The use of power supplies and a remote dose cord other than those listed in the electromagnetic emissions declaration may result in increased emissions or decreased immunity of the Pump.
|
Electromagnetic immunity declaration |
The Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Pump should assure that it is used in such an environment. |
Immunity test |
IEC 60601 test level |
Compliance level |
Electromagnetic environment – guidance |
Electrostatic discharge (ESD) IEC 61000-4-2 |
±8 kV contact ±15 kV air |
±8 kV contact ±15 kV air |
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. |
Electrical fast transient/burst IEC 61000-4-4 |
±2 kV for power supply lines ±1 kV for input/output lines |
±2 kV for power supply lines Not applicable |
Mains power quality should be that of a typical commercial or hospital environment. |
Surge IEC 61000-4-5 |
±1 kV differential mode ±2 kV common mode |
±1 kV differential mode ±2 kV common mode |
Mains power quality should be that of a typical commercial or hospital environment. |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 |
<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec |
<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec |
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Pump requires continued operation during power mains interruptions, it is recommended that the Pump be powered from an uninterruptible power supply or a battery. |
Power frequency 50/60 Hz) magnetic field IEC 61000-4-8 |
3 A/m |
400 A/m (IEC 60601-2-24) |
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
NOTE: UT is the a.c. mains voltage prior to application of the test level. |
Electromagnetic immunity declaration |
The Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Pump should assure that it is used in such an environment. |
Immunity test |
IEC 60601 test level |
Compliance level |
Electromagnetic environment – guidance |
|
|
|
Portable and mobile RF communications equipment should be used no closer to any part of the Pump, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. |
Conducted RF IEC 61000-4-6 |
3 Vrms 150 kHz to 80 MHz |
10 Vrms |
Recommended separation distance d=0.27*P1/2 |
Radiated RF IEC 61000-4-3 |
3 V/m 80 MHz to 2,5 GHz |
10 V/m |
d=0.27*P1/2 80MHz to 800 MHz d=0.54*P1/2 800MHz to 2,5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
 |
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. |
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. |
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Pump is used exceeds the applicable RF compliance level above, the Pump should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Pump. |
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m. |
Recommended separation distances between portable and mobile RF communications equipment and the Pump |
The Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pump as recommended below, according to the maximum output power of the communications equipment. |
Rated maximum output power of transmitter |
Separation distance according to frequency of transmitter m |
W |
150 kHz to 80 MHz d=0.27*P1/2 |
80 MHz to 800 MHz d=0.27*P1/2 |
800 MHz to 2,5 GHz d=0.54*P1/2 |
0.01 |
0.03 |
0.03 |
0.05 |
0.1 |
0.09 |
0.09 |
0.17 |
1 |
0.27 |
0.27 |
0.54 |
10 |
0.85 |
0.85 |
1.7 |
100 |
2.7 |
2.7 |
5.4 |
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. |
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. |
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. |
Safety Features and Fault Detection
Hardware Safety Features
Key hardware safety features include a watchdog timer circuit, motor driver and motor watchdog circuits, and a voltage detector circuit. Each safety circuit performs a unique function to insure the overall safety of the device.
Watchdog Timer Circuit
The microprocessor must send an appropriate signal to the watchdog circuit at least once per second. If the microprocessor does not, the watchdog circuit will time out and shut down the pump controller.
Watchdog timer circuitry is provided to monitor the status of the microprocessor and disable the motor and enable the audible alarm if the microprocessor fails to function properly. The microprocessor must strobe the watchdog circuit at least once every second in order to prevent the watchdog from performing its reset function. The reset output from the watchdog circuit is a pulse output. This acts to “jump startˮ the microprocessor. This unique feature allows the microprocessor to test the watchdog circuit on every power-up.
By setting a flag in the memory and not strobing the watchdog, the microprocessor can force a watchdog time-out. After being reset, the microprocessor checks the status flag to see if this was a time-out test. If so, the microprocessor continues normal power-up activities. If the reset occurred when the microprocessor was not expecting it, the microprocessor traps the event, sounds the audible alarm and displays an error message on the LCD.
Motor Driver/Motor Watchdog Circuit
Motor drive circuitry is composed of a series of power FET transistors, passive components, and 2 voltage comparators. Built into the motor drive circuitry is an RC timer which times how long the motor runs each time it is turned on. If the motor runs for more than an average of 3 seconds, the circuit will time out and disable the motor. A unique feature of this circuit is that control lines to and from the microprocessor circuit allow the microprocessor to perform a complete functional test of the motor drive circuit without running the motor. The microprocessor performs this test function every several minutes to assure its continued functionality. An input from the watchdog circuit prevents motor operation if the watchdog timer expires. The software verifies this function during the watchdog test described above.
Voltage Detector Circuit
Low voltage detection is performed by part of the watchdog circuit and by the microprocessor via software. Three low voltage levels are detected. The first 2 levels are detected by software and the third by hardware. The first level to be reached is the low battery warning threshold which occurs when the battery voltage decays to a nominal value of 2.4 volts when the motor is off or 1.8 volts when the motor is active. An analog to digital converter (ADC) built into the microprocessor allows the microprocessor, via software, to monitor the battery voltage. At the low battery warning threshold, the microprocessor enables a periodic series of beeps and displays a low battery warning message on the LCD. As the voltage operating the motor reaches a nominal value of 4.75 volts, the software disables delivery, places a battery depleted message on the LCD, and enables a constant two tone audible alarm. When the battery voltage decays to a nominal value of 1.0 volts, a hardware reset circuit is triggered which places the microprocessor in reset. This prevents ambiguous microprocessor operation when the battery voltage continues to decay. The hardware reset continues until the battery is completely discharged or until it is removed. Once the pump controller goes into low battery shutdown, only replacing the depleted batteries with new ones will clear the condition.
Software Safety Features
Hardware-related Software Safety Features
Program Memory Check
At power up and at regular intervals thereafter, the program memory is tested by calculating a cyclic redundancy code (CRC) on the program and then comparing it with the CRC stored with the program.
If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
RAM Memory Check
At power up, the random access memory is checked. A series of bit patterns is written to and read from each address in the RAM. If the read data is different from the written data, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Motor Circuit Check
At power up and at regular intervals thereafter, the motor circuit is checked to ensure that no power is being applied to the motor unless the motor is actually on. If the software detects power being applied to the motor at any other time, it will sound a continuous two-tone audible alarm and will no longer attempt to deliver medication. During every pump activation, the software checks to see whether the motor completes one activation. If the motor fails to turn, or fails to complete a cycle, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Keyboard Encoder Check
Every time the software receives data from the keyboard encoder, it is checked. If the data is not a valid key press, the software will disregard the key press. The keyboard is designed with redundant switches for

,

, and

. The software must detect that both switches are activated before taking any action.
Data Handling Software Safety Features
Data Stored in RAM
Before use, data associated with delivery and stored in RAM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Data Stored in EEPROM
Before use, data associated with delivery and stored in EEPROM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Data Stored in NOVRAM
Before use, data associated with delivery and stored in NOVRAM is tested by calculating a CRC on the data and then comparing it with the CRC stored with the data. If the stored and calculated CRCs do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Data Used in Calculations
Calculations on data used in some way to control the delivery of medication are performed redundantly.
The two calculated values are then compared. If the two values do not match, the software will display a system fault screen, turn on a continuous two-tone audible alarm, and stop all medication delivery.
Timer Data Registers
The data in the real time clock is checked at regular intervals. If the data is not reasonable, the software will turn on a continuous two-tone audible alarm and stop all medication delivery.
Annual Functional Inspection
Smiths Medical recommends annual functional inspections and tests on the CADD-Legacy® 1400 pump. Contact Smiths Medical to coordinate return and inspection of the pump.
CAUTION: CADD-Legacy® 1400 pumps are sealed units. A broken or damaged seal will, therefore, be considered conclusive evidence that the pump has been misused and/or altered, which voids any and all warranties. All service and repair of CADD-Legacy® 1400 pumps must be performed by Smiths Medical or its authorized agents. |
Collect Separately
This product contains electronic and other components (such as batteries) that may contain materials which, if disposed of with general household waste, could be damaging to the environment.
In accordance with Directive 2002/96/EC Waste Electrical and Electronic Equipment, Smiths Medical requires that residents of the European Union return this product for proper disposal at the end of its useful life.
If you are unsure of the proper disposal method, contact your local distributor for specific disposal instructions.
WARNING: There are potential health hazards associated with improper disposal of batteries, electronics, and contaminated (used) reservoirs and extension sets. Dispose of used batteries, reservoirs, extension sets and other used accessories, or a pump that has reached the end of its useful life, in an environmentally safe manner, and according to any regulations that may apply. |
Limited Warranty
Smiths Medical ASD, Inc. (the “Manufacturer”) warrants to the Original Purchaser that the infusion pump (the “Pump”), not including accessories, shall be free from defects in materials and workmanship under normal use, if used in accordance with this Operator’s Manual, for a period of one year from the actual date of sale to the Original Purchaser. THERE ARE NO OTHER WARRANTIES.
This warranty does not cover normal wear and tear and maintenance items, and specifically excludes batteries, administration sets, extension sets or any other accessory items or equipment used with the Pump.
Subject to the conditions of and upon compliance with this Limited Warranty, the Manufacturer will repair or replace at its option without charge (except for a minimal charge for postage and handling) any Pump (not including accessories) which is defective if a claim is made during such one-year period.
The following conditions, procedures, and limitations apply to the Manufacturer’s obligation under this warranty:
A. Parties Covered by this Warranty: This warranty extends only to the Original Purchaser of the Pump. This warranty does not extend to subsequent purchasers. The Original Purchaser may be a patient, medical personnel, a hospital, or institution which purchases the Pump for treatment of patients. The Original Purchaser should retain the invoice or sales receipt as proof as to the actual date of purchase.
B. Warranty Performance Procedure: Notice of the claimed defect must be made in writing or by telephone to the Manufacturer as follows: Smiths Medical ASD, Inc. 1265 Grey Fox Road, St. Paul MN 55112 USA, 1 800.258.5361 (USA). Notice to the Manufacturer must include date of purchase, model and serial number, and a description of the claimed defect in sufficient detail to allow the Manufacturer to determine and facilitate any repairs which may be necessary. AUTHORIZATION MUST BE OBTAINED PRIOR TO RETURNING THE PUMP. If authorized, the Pump must be properly and carefully packaged and returned to the Manufacturer, postage prepaid. Any loss or damage during shipment is at the risk of the sender.
C. Conditions of Warranty: The warranty is void if the Pump has been 1) repaired by someone other than the Manufacturer or its authorized agent; 2) altered so that its stability or reliability is affected; 3) misused; or, 4) damaged by negligence or accident. Misuse includes, but is not limited to, use not in compliance with the Operator’s Manual or use with nonapproved accessories. The Pump is a sealed unit, and the fact that the seal has been broken will be considered conclusive evidence that the Pump has been altered or misused. Removal or damage to the Pump’s serial number will invalidate this warranty.
D. Limitations and Exclusions: Repair or replacement of the Pump or any component part thereof is the EXCLUSIVE remedy offered by the Manufacturer. The following exclusions and limitations shall apply:
- No agent, representative, or employee of the Manufacturer has authority to bind the Manufacturer to any representation or warranty, expressed or implied.
- THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS OR USE OF THE PUMP FOR ANY PARTICULAR PURPOSE.
- The Pump can only be used under the supervision of medical personnel whose skill and judgment determine the suitability of the Pump for any particular medical treatment.
- All recommendations, information, and descriptive literature supplied by the Manufacturer or its agents are believed to be accurate and reliable, but do not constitute warranties.
E. Computer Program License:
- The Pump is intended to be used in conjunction with a particular Licensed Computer Program supplied by Manufacturer and use of any other program or unauthorized modification of a Licensed Computer Program shall void Manufacturer’s warranty as set forth above.
- The Original Purchaser and any users authorized by the Original Purchaser are hereby granted a nonexclusive, nontransferable license to use the Licensed Computer Program only in conjunction with the single Pump supplied by Manufacturer. The Licensed Computer Program is supplied only in machine-readable object code form and is based upon Manufacturer’s proprietary confidential information. No rights are granted under this license or otherwise to decompile, produce humanly readable copies of, reverse engineer, modify or create any derivative works based upon the Licensed Computer Program.
- All other terms and conditions of this Limited Warranty shall apply to the Licensed Computer Program.
The Manufacturer disclaims responsibility for the suitability of the Pump for any particular medical treatment or for any medical complications resulting from the use of the Pump. The Manufacturer shall not be responsible for any incidental damages or consequential damages to property, loss of profits, or loss of use caused by any defect or malfunction of the Pump.
This warranty gives the Original Purchaser specific legal rights, and the Original Purchaser may have other legal rights which may vary from state to state.
Index
A |
AC adapter |
accessory jack |
accuracy tests |
AC indicator light |
arrow keys |
B |
batteries, AA |
|
installing |
battery compartment |
battery contacts, cleaning |
battery, internal clock |
battery life |
biomed functions |
|
extra dose lockout |
|
morning dose lockout |
|
specifications |
|
upstream occlusion sensor on/off |
biomed functions code |
C |
cassette latch |
cleaning pump |
continuous rate |
|
programming |
D |
downstream occlusion sensor |
E |
ECG equipment, interference with |
electromagnetic emissions and immunity declarations |
extra dose |
|
Lockout |
|
programming |
|
starting |
|
stopping |
extra dose key |
G |
given |
|
clearing |
K |
keypad, keys |
L |
latch, cassette |
lock level |
|
changing |
lock level code |
M |
magnetic resonance imaging |
main screen |
medication cassette reservoir |
|
attaching |
|
removing |
morning dose |
|
Lockout |
|
starting |
|
stopping |
P |
power jack |
power-up |
priming |
programming |
programming, general |
programming screens |
R |
radiation, exposure to |
reservoir volume |
|
programming |
|
resetting |
S |
safety features |
|
hardware |
|
software |
security codes |
|
biomed functions code |
|
lock level code |
service due |
software version |
specifications |
|
biomed functions |
|
delivery |
|
general |
standards |
starting pump |
stopping pump |
symbols |
system definition |
T |
turning pump on/off |
U |
ultrasound, vii |
upstream occlusion sensor |
W |
warnings and cautions |
warranty |
Appendix A – Pump Programming Quick Reference for Healthcare Providers
This quick reference provides for step-by-step directions for several of the common pump programming tasks performed with the CADD-Legacy® 1400 pump. Additional pump information including warnings, cautions and more information on pump operations is located in the referenced sections of the pump Operator’s Manual. Please refer to the full prescribing information for DUOPA (carbidopa and levodopa) enteral suspension for indications and usage, contraindications, warnings, precautions, and adverse reactions.
Begin programming the pump by:
- Attaching a medication cassette reservoir
- Turning on the pump
For instruction on attaching a medication cassette reservoir, see Section 2, Pump Setup and Programming.
You will need the following items to complete these steps:
- Pump
- Medication cassette reservoir
Changing to Lock Level 0 (LL0)
Lock level 0 (LL0) allows the health care provider to adjust settings so they are appropriate for the patient. For more information on lock levels descriptions, see Section 1, General Description. For more information on changing the lock level, see Section 2, Pump Setup and Programming.
**Text Omitted**
Pump Programming Settings
Program the pump settings to customize the medication delivery inputs for the patient. For more information see Section 2, Pump Setup and Programming.
NOTE: Ensure that the pump is in lock level 0 (LL0) and

appears on the screen.
1. Press

Check for

on the screen.
NOTE: Not in Use is the default setting for the reservoir volume. The reservoir volume feature is not required for use, but is available at provider discretion.
2. Press

again. Check for

on the screen.
3. Press

or

to select the desired continuous rate.
4. Press

Pump Status: The continuous rate is now set.
5. Press

Check for

on the screen.
6. Press

or

to select the extra dose amount.
7. Press

Pump Status: The extra dose amount is now set.
8. Press

check for

on the screen
9. Press

if you wish to clear the amount given.
10. Press

NOTE: To change a setting again, press NEXT until the appropriate screen appears. Press

or

to adjust the setting, then press

to confirm.
Changing Lockout Times
Program the dose lockout times to customize medication delivery inputs for each patient. Lockout times will determine how often a patient can deliver a morning dose and an extra dose. These values should be determined during titration. For more information on DUOPA titration, refer to the full prescribing information for DUOPA (carbidopa and levodopa) enteral suspension.
The biomed functions allow the health care provider access to the extra dose lockout and morning dose lockout settings. For more information about biomed functions, see Section 4, Biomed Functions.
NOTE:
- Ensure that the pump is in lock level 0 (LL0) and

is on the screen.
- When entering a new lockout value, any lockout time previously in effect will be cleared.
**Text Omitted**
Programming the Morning Dose
Program the morning dose to customize medication delivery for the patient. For more information, see Section 2, Pump Setup and Programming.
CAUTION: Review programming screens when complete to make sure desired programming has been entered. Check to make sure unintended changes were not made to the morning dose, continuous rate, or extra dose volume. If unintended changes were made, go to the appropriate screen and program the desired value. |
1. The pump must be running with a cassette attached and in LL0 or LL1. Start the pump, if necessary and confirm the lock level settings.
NOTE: In LL0, programming in the full range is possible. In LL1, you can program up to the LL0 value.
2. Press


should appear on the screen.
3. Press

or

to select the desired morning dose volume.
4. Press

to store the morning dose volume.
Setting the Lock Level
For patient use, the pump must be set to lock level 2 (LL2) or lock level 1 (LL1). For more information on lock levels descriptions, see Section 1, General Description. For more information on changing the lock level, see Section 2, Pump Setup and Programming.
**Text Omitted**
Manufacturer:
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul, MN 55112 USA
Tel: 1 800 258 5361 (USA), +1 614 210 7300
www.smiths-medical.com

CADD, CADD-Legacy, and Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners.
© 2015 Smiths Medical. All rights reserved.
2015-01
40-6756-01C
